"BRYAN LABORATORY" "VERBAL WALTER" "6919 ALMEDA RD" "" "HOUSTON" "TX" "77021" "" "2/24/2009" "S/N 197606 - 150000 lbf SUPER L SR2 RETROFIT - w/ Machine Display - No Adjustments - Good Condition" "S/N 132885 - S-1000-2 EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "S/N 136066A - S-1000-2A EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "S/N 136070 - S-1000-2A EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "S/N 136295 - S-1000-2A EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "S/N 136066 - S-1000-1A EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "S/N 136069 - S-1000-8 EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "S/N 136067 - S-1000-1.4A EXTENS - w/ Machine # 197606 - No Adjustments - Good Condition" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "SERVICE AND CALIBRATION ON THE EQUIPMENT LISTED ABOVE" "" "" "" "" "" "" "" "" "" "" "" "" "" "Roger Cooper" "6/26/2009" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "" "REQUESTED TO CHECK SUPER-L LOAD CALIBRATION ALONG WITH S-1000-2 EXTENSOMETER. BOTH PIECES OF EQUIPMENT FOUND TO BE WELL WITHIN RATED ACCURACY REQUIRMENTS AND REQUIRED NO ADJUSTMENTS. THE SUPER-L WAS LESS THAT .5% ANYWHERE IN THE CALIBRATED RANGE AND THE EXTENSOMETER WAS B-1 CLASS THOUGH ITS FULL RANGE ON MOTION. CALIBRATION OF SYSTEM NOT CAUSING ANY PROBLEMS WITH TESTING OR RESULTS OBTAINED." "" "" "" "" "" "" "" "" "" "THIS I BELIEVE IS TO BE CHARGED TO LARRY WITH THE PIPE MILL NOT BRYAN LABS" "12:00 AM" "12:00 AM" "" "Roger Cooper"